Hacking the SEM: Crystal phase detection for nanoscale samples
25 Jan 2012
Custom modifications of equipment are an honoured tradition of the research lab. In a recent paper, two materials scientists at the National Institute of Standards and Technology (NIST) describe how a relatively simple mod of a standard scanning electron microscope (SEM) enables a roughly 10-fold improvement in its ability to measure the crystal structure of nanoparticles and extremely thin films.
By altering the sample position, they are able to determine crystal structure of particles as small as 10 nanometres. The technique, they say, should be applicable to a wide range of work, from crime scene forensics to environmental monitoring to process control in nanomanufacturing.
The technique is a new way of doing electron diffraction with an SEM. In standard SEM-based electron diffraction, the researcher analyses patterns that are formed by electrons that bounce back after striking atoms in the sample. If the sample is a crystalline material, with a regular pattern to the arrangement of atoms, these diffracted electrons form a pattern of lines that reveals the particular crystal structure or "phase" and orientation of the material.
The information, say NIST's Robert Keller and Roy Geiss, can be critical.
"A common example is titanium dioxide, which can exist in a couple of different crystal phases. That difference significantly affects how the material behaves chemically, how reactive it is. You need to add crystallographic identification to the chemical composition to completely characterise the material."
SEMs often are outfitted with an instrument for just this task, a device called an EBSD (electron back-scatter diffraction) detector. The problem, they say, is that below a certain size, the usual setup just doesn't work.
"You can determine the crystal structure of an isolated particle down to a size of about 100 to 120 nanometres, but below that the crystals are so small that you're getting information about the sample holder instead."
A somewhat more exotic instrument, the transmission electron microscope (TEM), does much better, they say, but samples below about 50 nanometres in size show very limited diffraction patterns because the higher-powered electron beam of the TEM just blasts through them.
The novel tweak developed by Keller and Geiss combines a little of each.
They moved the SEM sample holder closer to the beam source and adjust the angles so that instead of imaging electrons bouncing back from the sample, the "EBSD" detector is actually seeing electrons that scatter forward through the sample in a manner similar to a TEM. (They also came up with a unique method of holding samples to make this work.)